Customization: | Available |
---|---|
Warranty: | One Year |
Application: | Multifunction Clibrator |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features and Benefits
Summary of Source and Measure Functions
Function |
Measure |
Source |
DC V |
0~50V (the upper display ± 30V) |
0~10V |
DC mA |
0~24mA (the upper display ± 24mA) |
0~24mA |
Frequency |
1.000Hz ~ 100.00kHz |
0.00Hz~20.000kHz |
Resistance |
0~3200Ω |
0~3200Ω |
DC mV |
0~100mV (the upper display ± 200mV) |
0~100mV |
RTD |
Pt100, Pt1000, Cu50, Cu100 |
Pt100, Pt1000, Cu50, Cu100 |
TC |
E, J, K, T, B, R, S, N |
E, J, K, T, B, R, S, N |
Others |
Loop supply, Step, Ramp, Dual display |
Specifications
Specifications are based on a one year calibration cycle and apply from +18 ºC to +28ºC unless stated otherwise. All specifications assume a 10 minute warmup period.
DC Voltage Measurement
Range |
Maximum measurement range |
Resolution |
Accuracy ( % of reading + counts) |
|
ETX-1825 |
ETX-2025 |
|||
28V (upper display) |
-33V~33V |
0.001V |
0.05+2 |
0.025+2 |
200mV (upper display) |
-200mV~200mV |
0.01mV |
0.05+2 |
0.025+2 |
50V (lower display) |
-1V~60V |
0.001V |
0.05+2 |
0.02+2 |
100mV (lower display) |
-15mV~125mV |
0.01mV |
0.05+2 |
0.02+2 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. Input resistance: >1MΩ. |
DC mA measurement
Range |
Maximum measurement range |
Resolution |
Accuracy ( % of reading + counts) |
|
ETX-1825 |
ETX-2025 |
|||
20mA (upper display) |
-24mA~24mA |
0.001mA |
0.05+2 |
0.025+2 |
20mA (loop of upper display) |
0~24mA |
0.001mA |
0.05+2 |
0.025+2 |
20mA (lower display) |
0~24mA |
0.001mA |
0.05+2 |
0.02+2 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C.. Input resistance:<100Ω. |
Resistance Measurement (upper display)
Range | Maximum output range | Resolution | Accuracy ( % of reading + counts) | |
ETX-1825 | ETX-2025 | |||
400Ω | 0~440Ω | 0.1Ω | 0.05+2 | |
3200Ω | 420Ω~3300Ω | 1Ω | ||
On-off test | 0~200Ω | 1Ω | 0.05+1 | |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. | ||||
Maximum load voltage: 20V, equivalent to voltage of 20mA on 1000Ω load resistance. |
Resistance measurement (lower display)
Range | Maximum measurement range | Resolution | Accuracy(Ω) | |||
ETX-1825 | ETX-2025 | |||||
Two-wire, three-wire | Four-wire | Two-wire, three-wire | Four-wire | |||
400Ω | 0~440Ω | 0.01Ω | 0.25 | 0.15 | 0.15 | 0.1 |
3200Ω | 420Ω~3600Ω | 0.1Ω | 1.5 | 1 | 1 | 0.5 |
-10°C~18°C,+28°C~55°C temperature coefficient,±0.005%FS/°C. | ||||||
Exciting current during measurement: 400Ω:0.4mA±10%;3200Ω:0.2mA±10%. | ||||||
Two-wire: Does not include lead resistance. | ||||||
Three-wire: Assumes matched leads with a total resistance not exceeding 25Ω. |